Technical specification |
 |
 |
| |
Measuring principle |
|
Scattered light forward |
|
 |
 |
 |
| |
Measuring ranges |
|
Dust content: 0 ... 5 mg/m³ / 0 ... 200 mg/m³;
higher concentrations on request |
|
 |
 |
 |
| |
Process temperature |
|
-25 ... 300 °C (-13 ... 572 °F), non-condensing, maximum temperature depends on type |
|
 |
 |
 |
| |
Process pressure |
|
-50 ... 2 hPa;
other pressure ranges with external purge-air unit on request. |
|
 |
 |
 |
| |
Duct diameter |
|
Min.: 0.5 m (19.7 in) |
|
 |
 |
 |
| |
Ambient temperature |
|
-25 ... 60 °C (-13 ... 140 °F),
max. +45 °C (113 °F) for MCU with purge-air unit |
|
 |
 |
 |
| |
Ambient humidity |
|
0 ... 95 %, relative humidity |
|
 |
 |
 |
| |
Conformities |
|
2000/76/EC, 2001/80/EC, GOST, MCERTS, German Clean Air Regulations (TA-Luft), U.S. EPA compliant |
|
 |
 |
 |
| |
Electrical safety |
|
CE |
|
 |
 |
 |
| |
Enclosure rating |
|
IP 66 |
|
 |
 |
 |
| |
Analogue outputs |
|
3 outputs:
0/2/4 ... 20 mA, 750 Ohm;
further analogue outputs when using I/O modules |
|
 |
 |
 |
| |
Analogue inputs |
|
2 inputs:
0 ... 20 mA, 0 ... 5/10 V;
further analogue inputs when using I/O modules |
|
 |
 |
 |
| |
Digital outputs |
|
5 relay contacts:
120 V AC, 1 A (30 V DC, 2 A);
further digital outputs when using I/O modules |
|
 |
 |
 |
| |
Digital inputs |
|
4 inputs;
further digital inputs when using I/O modules |
|
 |
 |
 |
| |
Interfaces |
|
RS-232, USB 1.1 |
|
 |
 |
 |
| |
Device type |
|
Measuring probe type |
|
 |
 |
 |
| |
Operation |
|
Via SOPAS software or LC display (option) |
|
 |
 |
 |
| |
System components |
|
Sender/receiver unit, control unit MCU |
|
 |
 |
 |
| |
Test functions |
|
Automatic check of zero and reference point, contamination monitoring |
|
 |
 |
 |
| |
Bus protocol |
|
Ethernet (option)
PROFIBUS-DP (option) |
|
 |
 |
 |
| |
|
|
|
|
 |