SHC500, Gravimat
|
|
| |
Mobile dust measuring system For control and gravimetric comparison measurements on dust concentration monitors. A partial gas flow is sucked off isokinetically via a filter head probe (control in real time). The dust content ist determined by weighting the dust collector mass before and after extraction.
|
|
 |
 |
 |
 |
 |
Technical specification |
 |
 |
| |
Measuring values |
|
Dust concentration in mg/m³ (operation state, standard state) |
|
 |
 |
 |
| |
Measuring range |
|
0 ... 50,000 mg/m³ (0 ... 21.85 gr/ft³) |
|
 |
 |
 |
| |
Measurement principle |
|
Gravimetry |
|
 |
 |
 |
| |
Compliance |
|
VDI Guideline 2066, ISO 9096, EN 13284-1, EPA method 17 |
|
 |
 |
 |
| |
Protection category |
|
IP 54 (automatic unit open; closed IP 65) |
|
 |
 |
 |
| |
Interface |
|
RS232 |
|
 |
 |
 |
| |
System modules |
|
Filter head probe, case with set of dust collectors, automatic unit, (laptop) |
|
 |
 |
 |
| |
Test function |
|
Self-testing function |
|
 |
 |
 |
| |
Remarks |
|
- Compact measuring system, easy to transport
- No dust loss at the handling by patented dust sampling system. Allows high measuring precision also at low dust concentrations.
- Automatic test value recording, system control and measurement storage.
- Real time isokinetic control
- Measurement results are immediately available after dust sampling (type dependent also automatic)
|
|
 |
 |
 |
| |
|
|
|
|
 |
|
 |
|