Search   Start
 
GM901
 
 
Gas analyzer for CO
In addition to its modular concept – the Evaluation Unit can be deployed up to 1000 m from the measuring point – the GM901 CO Analyzer is available as a cross-duct or probe version. As a result, the GM901 is ideally suited to a broad range of application, even for difficult measuring tasks (e.g. high dust concentrations, overpressure, „wet“ flue gases, or extremely high measuring gas concentrations).
Applications
Process control, e.g. for monitoring fuel-burning plants in power plants, cement industry, steel industry or chemical industry.
 
Technical specification
  Measuring component   CO  
  Measuring range   500 ... 20,000 ppm x m  
  Measuring path   0.5 ... 8 m (1.6 ... 26 ft), cross-duct  
  Measurement principle   Gas filter correlation (IR)  
  Measuring gas temperature   Cross-stack: 250 °C (482 °F) standard; 370 °C (698 °F) max.
Probe version: 220 °C (428 °F), permanent operation;
260 °C (55 °F), no perm. operation, others on request
 
  Compliance   TÜV approved, U.S. EPA compliant  
  Protection category   IP 65  
  Measured value output   0/2/4 ... 20 mA analog output  
  Interfaces   RS232, RS422/485, CAN-Bus, optional PROFIBUS  
  Status / control signals   Relay outputs for status signals  
  System modules   Transmitter, receiver and evaluation unit, measuring probe  
  Test function   Manual test function  
  Remarks  
  • Cross-stack and probe version available
  • Easy to install, commission, and project plan
  • Verifiable using test gases
  • User-friendly operation from the evaluation unit
  • Extremely low maintenance requirements
 
     