Search   Start
 
MONOCOLOR1 N and MONOCOLOR2 Ex
 
 
Hydrogen sulfide gas analyzer
The H2S gas analyzer MONOCOLOR operates semi-continuously by the colorimetric measuring principle with a dry reaction on a reagent paper strip. The intensity of the reagent paper stain is proportional to the mean concentration of H2S and to the flowrate.This measuring principle is very selective, even in case of different sulfur compounds in the sample gas and allows the measurement of small H2S concentrations.
Applications
The MONOCOLOR is suitable for monitoring of hydrogen sulfide in natural gas, bio gas, landfill gas, coke oven gas, for ambient air and emission monitoring in desulfurization plants, the pulp & paper, chemical and petrochemical industry.
 
Technical specification
  Measured component   H2S  
  Measuring range   0 ... 0.8 ppm up to
0 ... 1.6 vol.%
 
  Measurement principle   Colorimetric with lead acetate or silver nitrate paper  
  Measuring gas pressure   60 …120 hPa (0.87 ... 1.7 PSI)  
  Measuring gas temperature   5 … 45 °C (41 ... 113 °F)  
  Ex protection   MONOCOLOR2 Ex: EEx de (ib) IIB T4, PTB 02 ATEX 2177 X  
  Protection category   MONOCOLOR1 N: IP 20
MONOCOLOR2 Ex: IP 65
 
  Measured value qutput   MONOCOLOR1 N: analog, 0, 2, 4 ... 20 mA
MONOCOLOR2 Ex: analog, 4 ... 20 mA, intrinsically safe
 
     
Shopping Cart