ModularSystem S700
|
|
| |
Gas analysis with system The S700 modular system makes configuring tailored applications particularly easy. An S700 housing can be equipped with up to 3 analyzer modules for compact and cost-effective solutions. A total of 6 different analyzer modules are available for analyzing up to 60 gas components. Depending on the measuring task, installation site, and ambient conditions:
- S710 (19” housing) for standard applications
- S715 for rough ambient conditions or explosion protection zone 2
- S720 Ex for industrial use in explosion protection zone 1
|
|
 |
 |
 |
 |
 |
Technical specification |
 |
 |
| |
Measured components |
|
up to 60 components, e.g. CO, CO2, SO2, NO, NH3, H2O, N2O, CH4, C2H2, C2H4, C2, H6, C3H6, C3H8, C4H6, C4H10, C6H14, O2 |
|
 |
 |
 |
| |
Measuring range |
|
depends on application |
|
 |
 |
 |
| |
Measurement principle |
|
NDIR, thermal conductivity, interference filter correlation, rotating diamagnetic dumbbell, electrochemical cell |
|
 |
 |
 |
| |
Measuring gas pressure |
|
16 ... 250 hPa (0.23 ... 3.6 PSI) |
|
 |
 |
 |
| |
Measuring gas temperature |
|
0 ... 45 °C (32 ... 113 °F) |
|
 |
 |
 |
| |
Compliance |
|
cCSAus; CE; TÜV, German Clean Air Regulations, Federal German Pollution Control Act (13th/17th/27th Implementing Ordinance) |
|
 |
 |
 |
| |
Ex protection |
|
S715: II 3 G EEx n R P II T6 (Class I, Division 2, Groups A, B, C and D)
S720: II 2 G EEx d ia IIC T6 |
|
 |
 |
 |
| |
Protection category |
|
S710: IP 20
S715: IP 65 (Nema 4X)
S720 Ex: IP 65 (Nema 7) |
|
 |
 |
 |
| |
Interface |
|
RS232 |
|
 |
 |
 |
| |
Remarks |
|
- UNOR - NDIR for one component
- MULTOR - NDIR up to 3 components
- THERMOR - termal conductivity
- FINOR - interference filter correlation - up to 3 components
- OXOR-E - oxygen analyzer (electrochemical)
- OXOR-P - oxygen analyzer (paramegnetic)
|
|
 |
 |
 |
| |
|
|
|
|
 |
|
 |
|