Extractive Online Analyzer for Measurement of Organic Water Pollutants (TOC)
The TOC-analyzer TOCOR oxidizes the organic carbon in the water sample to CO2 which is then measured by a NDIR gas analyzer. Previously, anorganic carbon is removed by a stripper. The required carrier gas is produced by the TOCOR700 itself, so it can be operated indepentendly from instrument air. Two variants of the TOCOR700 are avalilable:
  • TOCOR700 TH, with a thermal high-temperature reactor, is especially suitable as a precise analyzer device for unknown water contaminants.
  • TOCOR700 UV, with its UV reactor can be used for monitoring of limits or in case of known water contaminants.
Applications
TOC-monitoring in waste water, cooling, surface, river and trinking water. For environmental monitoring, industrial or municipal waste water treatment plants, in power stations, in chemical and petrochemical industry.
 
Technical specification
  Measured components   TOC (incl. VOC), TC  
  Measuring range   0 ... 2 mg/l C or higher  
  Measuring principle   Thermal-catalytical or UV oxidation to CO2 and measurement by a NDIR gas analyzer  
  Water consumption   max. 200 ml/h  
  Ambient temperature   +5 ... +45 °C (41 ... 113 °F)  
  Measured value output   0/4 ... 20 mA, linear, electrically isolated, max. load: 500 Ohm; max. 4 outputs  
  Status signals   2, for failure and maintenance, potential-free contacts,
max. load: 48 V / 0,5 A
 
  Dimensions
(W x H x D)
  depending on device design  
  Remarks   Optional available:
  • Variants for application in hazardous areas Zone 1 or 2
  • Cabinet: stand-alone or wall-mounting
  • Freezing protection or air conditioning
 
     